As per the manufacturer's evaluation report: the neutral electrode of the cutting loop is broken off, insulation is melted, the cutting wire melted off; the break of the ceramic tip, as well as the welding point on the endoscope, are collateral damages following the damage of the cutting loop.The remaining electrode shows bending, and the resulting heat cracked the ceramic tip.
|
As per a report we received from the factory in (b)(6) in regard to an incident that occurred in (b)(6): when surgeon was performing a tcr, at around three minutes, they found the cutting loop burning and the scope was damaged and the white ceramic of sheath also broken.An x-ray confirmed everything was retrieved; there was no harm to patient.
|