• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. HEALON PRO; OVDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. HEALON PRO; OVDS Back to Search Results
Model Number TH85ML
Device Problem Contamination (1120)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/26/2020
Event Type  malfunction  
Manufacturer Narrative
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.If implanted; give date: n/a.Healon pro is not an implantable device.If explanted; give date: n/a.Healon pro is not an implantable device; therefore, not explanted.Initial reporter phone: (b)(6).Device evaluation: the healon pro unit was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be performed.Manufacturing record evaluation: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification. a search revealed that no additional complaints for this order number have been received.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that some fibers were observed in the patient eye while inserting the healon pro.The surgeon tried to remove the fiber with tweezers but it did not succeed.Surgeon then flushed the eye with balance saline solution (bss) to remove the fiber.However, it is unknown if the fiber was removed.No additional information was provided.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALON PRO
Type of Device
OVDS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10318368
MDR Text Key219528256
Report Number3004750704-2020-00034
Device Sequence Number1
Product Code LZP
UDI-Public(01)(17)230131(10)UH31072
Combination Product (y/n)Y
PMA/PMN Number
P810031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 01/01/2005,10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberTH85ML
Device Catalogue Number10271012
Device Lot NumberUH31072
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received10/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-