The device was not returned for evaluation; it was discarded at the hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.No corrective actions will be taken at this time.Lot number was not provided, therefore review of the manufacturing records could not be completed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications are well described in the literature.With any hemodynamic monitoring, pressure readings can change quickly and dramatically.Pressure tubing that is used with swan ganz catheters can also be a contributing factor to inaccurate values.In regard to the pressure tubing used with swan ganz catheters, it should be noted that poor dynamic response can be caused by air bubbles, clotting, excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.Pressure readings should correlate with the patient¿s clinical manifestations.It is unknown whether user or procedural factors contributed to the stated event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|