Model Number M3-0A |
Device Problem
Fogging (1253)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.The device was visually inspected and the outer tube was found to be bent and there was excessive debris underneath the glass cover on the distal tip.If additional information becomes available at a later time, this report will be supplemented.
|
|
Event Description
|
It was reported that the telescope image was foggy.There was no patient injury reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information (h6, h10) regarding the reported event.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.
|
|
Manufacturer Narrative
|
Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.
|
|