• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404232
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem Scarring (2061)
Event Date 07/06/2020
Event Type  Injury  
Event Description
It was reported that the left cylinder was "open inside the corpora cavernosa" and the patient had his inflatable penile prosthesis (ipp) cylinders and pump removed and replaced.It was explained that when the left cylinder was explanted, it was very difficult to pull it out.An open cut was revealed and seemed to be broken in layers with blood inside the cylinder and fibrosis was identified across the corpora cavernosa.It was added that there was no saline solution inside the device including the reservoir.The device was not working at all.The patient was stable after this surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
10500 bren road w
minnetonka MN 55343
Manufacturer Contact
alyson harris
10700 bren road w
minnetonka, MN 55343
4089353452
MDR Report Key10355530
MDR Text Key201340582
Report Number2183959-2020-03284
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953003450
UDI-Public00878953003450
Combination Product (y/n)N
Reporter Country CodeRQ
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/24/2016
Device Model Number72404232
Device Catalogue Number72404232
Device Lot Number888562005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/04/2014
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
-
-