Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The device was received.The reported lot number was not confirmed as the packaging was not returned.During the visual inspection, the stent device was returned within the catheter.There were no anomalies noted on the catheter.The stent delivery wire was kinked bent distally and proximally.The stent had deployed and was not returned.Functional testing not done as the stent was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The stent delivery wire was seen to be kinked bent and the stent was not returned and presumed deployed.The as reported event can be confirmed based on the analysis.The as reported and as analyzed stent delivery wire kinked bent and stent deployed prematurely during use will be assigned procedural factors this complaint appears to be associated with a product that met stryker and design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.
|