• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM CATRX ASPIRATION CATHETER; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM CATRX ASPIRATION CATHETER; DXE Back to Search Results
Model Number CATRXKIT
Device Problem Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/15/2020
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure in the right coronary artery (rca) using an indigo system catrx aspiration catheter (catrx), a non-penumbra sheath, a non-penumbra guide catheter, and two non-penumbra balloon catheters.During the procedure, the physician used the two balloon catheters to perform balloon angioplasty and open the very stenosed vessel.Next, while attempting to advance the catrx passed the thrombotic lesion in order to aspirate, the catrx would not advance; therefore, it was removed.The procedure was completed using two overlapping drug-eluting stents.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM CATRX ASPIRATION CATHETER
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key10374128
MDR Text Key201891039
Report Number3005168196-2020-01112
Device Sequence Number1
Product Code DXE
UDI-Device Identifier00814548017556
UDI-Public00814548017556
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K163618
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,07/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/12/2023
Device Model NumberCATRXKIT
Device Catalogue NumberCATRXKIT
Device Lot NumberF96907
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received07/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/12/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
-
-