The alleged product was an unknown prismaflex set.The actual device was not available; however, a photograph of the sample was provided for evaluation.The visual inspection of the provided picture confirmed that the cone of the male luer lock of the effluent line was cracked and stuck in the female luer lock of the effluent bag.The reported condition was verified.The cause was manufacturing related.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
It was reported that during treatment with a prismaflex set, the luer connector of the effluent bag was occluded with a plastic piece, which did not allow total amount of effluent into the bag.Small amounts of leakage were visible outside of the bag.Upon changing the effluent bag, large amounts of effluent came from effluent line.There was no patient injury or medical intervention associated with this event.
|