• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 13X111MM 111MM T1; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 13X111MM 111MM T1; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 07/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: foreign country: (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-03083, 0001825034-2020-03084, 0001825034-2020-03085, 0001825034-2020-03086, 0001825034-2020-03087, 0001825034-2020-03089.
 
Event Description
It was reported the warehouse investigated circulated items in their stock and identified debris in the sterile package.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Evaluation of the returned product/photographs provided confirmed there is debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Sterility has not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage.This device falls within the scope of a previous corrective action, the purpose of which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.As part of this, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.Upon investigation, it has been determined that the debris in the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR MP T1 PPS 13X111MM 111MM T1
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10383202
MDR Text Key202193261
Report Number0001825034-2020-03089
Device Sequence Number1
Product Code OQG
UDI-Device Identifier00880304517325
UDI-Public(01)00880304517325
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-145130
Device Lot Number6047653
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-