NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number CL-325-10
Device Problems Break (1069); Impedance Problem (2950)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/13/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The explanted product was not returned to neuropace for investigation.Review of the lead impedance and ecog data was performed and the data is consistent with a lead break.Unable to determine root cause of the lead break.
 
Event Description
During a review of rns neurostimulator data by the treating clinician, artifact was observed on the ecog recorded from the left strip lead , indicating a potential lead break.After discussion with the patient, it was determined that this signal change occurred following a fall due to a seizure, consistent with the patient's pre-implant seizure semiology.The patient was subsequently scheduled for a lead replacement procedure, which was performed without complication.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view, ca
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view, ca
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, ca 
2382788
MDR Report Key10387984
MDR Text Key202891742
Report Number3004426659-2020-00026
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005069
UDI-Public010085554700506917210123
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCL-325-10
Device Catalogue Number1007609
Device Lot Number25330-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age55 YR
-
-