Model Number CI-1400-01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Necrosis (1971); Staphylococcus Aureus (2058); Pressure Sores (2326)
Event Date 07/16/2020
Event Type  Injury  
Event Description
The recipient is reportedly experiencing a staphylococcus aureus and proteus mirabilis infection at the magnet site.The recipient is presenting with pressure ulcers and skin necrosis at the right of the implant site.The cause of the pressure ulcers is unknown.The recipient's device was explanted.The recipient will not be reimplanted at this time.The recipient is healing.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Additional information: the recipient's infection has resolved.No further medical intervention was needed.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10398617
MDR Text Key202622634
Report Number3006556115-2020-00796
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 07/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/01/2012
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-