• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 0DEG 30MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL / POLYMER, NON-POROUS, CALC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH LRG TAP PRI MOD NCK 0DEG 30MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL / POLYMER, NON-POROUS, CALC Back to Search Results
Catalog Number NLS-300000B
Device Problems Device-Device Incompatibility (2919); Insufficient Information (3190)
Patient Problems Inflammation (1932); Injury (2348); Reaction (2414)
Event Date 05/03/2013
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There has been another event for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported elevated levels of cobalt and chromium revealed in bloodwork is considered to be under the scope of this recall.No further investigation is required.Device not returned.
 
Event Description
It was reported that the patient is asymptomatic.Additional information received from legal on 7/22/2020: plaintiff was implanted with a rejuvenate modular hip stem on his left hip on or about (b)(6) 2011.It is further alleged that the plaintiff had the left hip stem at issue explanted on (b)(6) 2019 due to elevated levels of cobalt an chromium revealed in bloodwork.
 
Manufacturer Narrative
An event regarding elevated metal ions involving a rejuvenate modular device was reported.The event was confirmed.Method & results: -device evaluation and results: device evaluation was not performed as no devices were received.-device history review: review of device history records found the devices in the reported lot were accepted into final stock with no reported discrepancies.-complaint history review: the complaint databases show there have been other events for the reported lot.Similar events have occurred for the catalog number and product family.These events were determined to be associated with ra 2012-067.Conclusions voluntary recall ra 2012-067 was initiated for abgii and rejuvenate modular stems and necks due to the potential risks associated with these devices.The reported elevated metal ions is considered to be under the scope of this recall.No further investigation is required.
 
Event Description
It was reported that the patient is asymptomatic.Additional information received from legal on 7/22/2020: plaintiff was implanted with a rejuvenate modular hip stem on his left hip on or about (b)(6) 2011.It is further alleged that the plaintiff had the left hip stem at issue explanted on (b)(6) 2019 due to elevated levels of cobalt an chromium revealed in bloodwork.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LRG TAP PRI MOD NCK 0DEG 30MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL / POLYMER, NON-POROUS, CALC
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10401637
MDR Text Key202722589
Report Number0002249697-2020-01662
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
PMA/PMN Number
K092561
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Remedial Action Recall
Type of Report Initial,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Catalogue NumberNLS-300000B
Device Lot Number36688602
Was Device Available for Evaluation? No
Date Manufacturer Received08/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2090-2012
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-