According to the reporter, during a laparoscopic sigmoidectomy, when squeezing the handle, a crack breaking sound was heard, and the clip became unable to be loaded.Another device was used to complete the case.There was no patient injury.
|
Additional information: g4, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the handle was broken and out position, in a manner consistent of cycling through the safety interlock.Functional testing observed that the jaws would not close upon actuation of the handle.The shaft was removed from the instrument body and the body was disassembled for visualization of internal components.The safety interlock channel lugs were broken.It was reported that the clips did not load properly into the jaws as expected and the device gave unexpected or uncharacteristic audible feedback of normal use.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when an attempt is made to fire the instrument without first loading a clip by using the trigger.The safety interlock feature is designed to prevent the jaws from closing on a vessel in the absence of a loaded clip.If an attempt is made to forcibly fire the instrument while engaged in interlock, the lugs are designed to break as noted and the interlock feature continues to function as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|