Model Number 176615
Device Problems Mechanics Altered (2984); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a laparoscopic sigmoidectomy, when squeezing the handle, a crack breaking sound was heard, and the clip became unable to be loaded.Another device was used to complete the case.There was no patient injury.
 
Manufacturer Narrative
Additional information: g4, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the handle was broken and out position, in a manner consistent of cycling through the safety interlock.Functional testing observed that the jaws would not close upon actuation of the handle.The shaft was removed from the instrument body and the body was disassembled for visualization of internal components.The safety interlock channel lugs were broken.It was reported that the clips did not load properly into the jaws as expected and the device gave unexpected or uncharacteristic audible feedback of normal use.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when an attempt is made to fire the instrument without first loading a clip by using the trigger.The safety interlock feature is designed to prevent the jaws from closing on a vessel in the absence of a loaded clip.If an attempt is made to forcibly fire the instrument while engaged in interlock, the lugs are designed to break as noted and the interlock feature continues to function as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key10402781
MDR Text Key202755711
Report Number9612501-2020-01154
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521057807
UDI-Public10884521057807
Combination Product (y/n)N
PMA/PMN Number
K143644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number176615
Device Catalogue Number176615
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2020
Date Manufacturer Received09/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-