VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Information (3190)
Event Date 07/20/2020
Event Type  malfunction  
Manufacturer Narrative
A vyaire field service representative (fsr) went onsite and evaluated the ventilator.It was found that the delta p is out of specification during the performance check.According to fsr the unit is out of calibration.The fsr calibrated the airway pressure module and also performed a ddi calibration.Performed the driver controller and power width module calibration.After all calibrations, performed another operational verification procedure (ovp).Additionally, fsr found the airway pressure hose was almost disconnected internally.
 
Event Description
The customer reported that map was fluctuating 3100a device.At this time, there is no information regarding patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
peter sparacio
26125 n. riverwoods blvd.
mettawa, il 
3273284
MDR Report Key10411439
MDR Text Key203122055
Report Number2021710-2020-12400
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20000225
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/25/2000
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-