BAXTER HEALTHCARE CORPORATION CAPD DISCONNECT Y SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number JIC8328
Device Problem Misconnection (1399)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a uv flash disconnect y-set had a connection issue.This was further described as, while connecting the y-set to a transfer set, the spike of the transfer set bent.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: d10, h3, h6.H10: one actual sample was received for evaluation.A visual inspection with the naked eye noted a damaged molded port on the drain bag.The reported condition was verified.The cause of the damage could not be determined; however, a probable cause was a miss-spike during flash machine connection causing the damage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPD DISCONNECT Y SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10417086
MDR Text Key203281302
Report Number1416980-2020-05000
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier04987456383214
UDI-Public(01)04987456383214
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/22/2021
Device Catalogue NumberJIC8328
Device Lot NumberH16H22066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2020
Date Manufacturer Received09/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UV FLASH TRANSFER SET.
-
-