BAXTER HEALTHCARE CORPORATION CAPD DISCONNECT Y SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number JIC8328 |
Device Problem
Misconnection (1399)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a uv flash disconnect y-set had a connection issue.This was further described as, while connecting the y-set to a transfer set, the spike of the transfer set bent.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information: d10, h3, h6.H10: one actual sample was received for evaluation.A visual inspection with the naked eye noted a damaged molded port on the drain bag.The reported condition was verified.The cause of the damage could not be determined; however, a probable cause was a miss-spike during flash machine connection causing the damage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|