VDW GMBH VDW SILVER RECIPROCATING MOTOR; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number V041153000000
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where stalling has caused file separation.Since separation of a file could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Motor connection cable (loose contact) defective, battery (voltage breaks down under load) defective.Micromotor sm15291 (detected oil in motor) defective, contra angle 99927 (2016) and foot control 182612 do not indicate a defect.(st: 47; fi: 40; working hours: 58:22:30; enough is: 30).Please note that while this product is not sold in the us, it is considered similar to products that are marketed in the us by dentsply sirona.Note: the covid-19 pandemic caused a disruption in normal business activities, resulting in late submission of this report.
 
Event Description
In this event it was reported that a silver reciproc motor stops during use; no injury resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VDW SILVER RECIPROCATING MOTOR
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern, 81737
GM  81737
Manufacturer (Section G)
ADVANCED TECHNOLOGY RESEARCH (A.T.R.) S.R.L.
via san donato 1
pistoia, 51100
IT   51100
Manufacturer Contact
karl nittinger
221 w. philadelphia st.
suite 60w
york, pa 
8494424
MDR Report Key10431847
MDR Text Key203699488
Report Number9611053-2020-00125
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 08/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberV041153000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/16/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2020
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-