Catalog Number 22-4038 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/30/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during use the firstpass suture passer, self capture had passed x3 sutures from a 2.8mm q-fix implant, when the jaw became locked onto the tissue, eventually it was released, the practitioner decided to open another firstpass st; the procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
H10: h2, h6.The device used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined.A review of the device history records for the device showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Without the reported product a visual and functional evaluation cannot be performed and customer¿s complaint cannot be confirmed.Factors that could have contributed to the reported event include: (1) excessive force (2) tissue thickness.(3) damage or debris in the device tip between passes.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
|