BAXTER HEALTHCARE CORPORATION CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Catalog Number 101025 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/23/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during treatment with a gambro cartridge set, an external blood leak was observed between the venous line connection and the venous fistula.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Additional information: h3, h6 and h10.H10: the actual device was not available; however a reference picture was provided for evaluation.The visual inspection revealed a detached blue luer collar resulting in a leak.The reported condition was verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|