The reported event was inconclusive.No sample was returned for evaluation and unable to determine root cause of reported problem.Potential root cause for this failure mode could be user related (example: over aspirated, incorrect syringe)/collapse lumen/sac close eye/valve damage.The lot number was unknown; therefore, the device history record could not be reviewed.Unable to perform labelling review due to unknown product code.Although the product family was unknown, the latex foley catheter product ifus were found to be adequate based on past reviews.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|