U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD UNKNOWN BIRMINGHAM HIP RESURFACING (BHR) CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD UNKNOWN BIRMINGHAM HIP RESURFACING (BHR) CUP; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120160
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Patient Problems Pain (1994); Scar Tissue (2060); Synovitis (2094); Metal Related Pathology (4530)
Event Date 10/16/2019
Event Type  Injury  
Event Description
It was reported that, after a bhr tha construct had been implanted on the right hip, the patient experienced persistent pain, synovitis and metallosis.A revision surgery was performed to treat the adverse event.The patient outcome is unknown.
 
Manufacturer Narrative
H3, h6: it was reported that a revision surgery was performed on the right hip.The devices involved were all used in treatment.As of today additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, hemi head, stem and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.Similar complaints have been identified for sleeve and this failure will continue to be monitored.No other similar complaints were identified for the head, cup and stem.In the absence of the actual devices, the production records were reviewed for the known devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.The revision document does not note findings consistent with the complaint description of metallosis.The reported pain may be associated with the scar tissue; however, the root cause for the scar tissue cannot be concluded but some patients can be genetically predisposed as it is a known complication of joint surgeries and is related to the procedure and not the device.With the information provided the root cause of the excessive amount of fluid and mild synovitis cannot be confirmed and it cannot be concluded that the reported reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN BIRMINGHAM HIP RESURFACING (BHR) CUP
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa warwickshire
UK 
MDR Report Key10438914
MDR Text Key203978610
Report Number3005975929-2020-00302
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number74120160
Device Lot Number086149
Was Device Available for Evaluation? No
Date Manufacturer Received03/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-