|
Model Number N/A |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-03318, 0001825034-2020-03319, 0001825034-2020-03320, 0001825034-2020-03321, 0001825034-2020-03322, 0001825034-2020-03323.
|
|
Event Description
|
It was reported the circulated items in the warehouse identified debris in sterile packages.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d10; g4; h2; h3; h6 -the event reports that debris was identified inside the sterile packaging.-this was identified at a zb distribution centre, therefore there was no patient, user, or stakeholder harm.-evaluation of the returned product/photographs provided confirmed there is debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Sterility has not been compromised.The reported event is confirmed.-the likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage.-this device falls within the scope, the purpose of which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.The pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.Upon reassessment of the reported event, sterility has not been compromised and was determined to be not reportable.The initial report was forwarded in error and should be voided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d1, d2, d4, d10, h2, h3.Corrected: d4 catalog number.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Search Alerts/Recalls
|
|
|