• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR T1 PPS 15X150MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR T1 PPS 15X150MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 07/30/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-03318, 0001825034-2020-03319, 0001825034-2020-03320, 0001825034-2020-03321, 0001825034-2020-03322, 0001825034-2020-03323.
 
Event Description
It was reported the circulated items in the warehouse identified debris in sterile packages.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d10; g4; h2; h3; h6 -the event reports that debris was identified inside the sterile packaging.-this was identified at a zb distribution centre, therefore there was no patient, user, or stakeholder harm.-evaluation of the returned product/photographs provided confirmed there is debris inside the sterile packaging which is consistent with the appearance of foam debris from the foam packaging inside the sterile barrier.Sterility has not been compromised.The reported event is confirmed.-the likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage.-this device falls within the scope, the purpose of which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.The pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.Upon reassessment of the reported event, sterility has not been compromised and was determined to be not reportable.The initial report was forwarded in error and should be voided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR T1 PPS 15X150MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10441976
MDR Text Key204009241
Report Number0001825034-2020-03321
Device Sequence Number1
Product Code OQG
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-105150
Device Lot NumberR3305275A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-