ERIKA DE REYNOSA, S.A. DE C.V. COMBI SET; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Model Number 03-2622-3 |
Device Problems
Fluid/Blood Leak (1250); Separation Problem (4043)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
|
|
Event Description
|
A user facility biomedical technician (biomed) reported that a bloodline leak occurred at an unknown phase of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an external blood leak.The leak was visually observed from where the heparin line connects to the rest of the main bloodline tubing.The patient¿s estimated blood loss (ebl) was between 30 ml to 150 ml, however the exact amount was unknown.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on the same machine and treatment completed successfully with new supplies.The complaint device was reported to not be available to be returned to the manufacturer for evaluation because it was discarded.
|
|
Manufacturer Narrative
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the customer for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius combiset bloodlines shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
|
Search Alerts/Recalls
|
|
|