Unique device identifier (udi) : (b)(4).The valve was returned for evaluation.Device history record (dhr) - the lot history record was reviewed for completeness during the release process to inventory.At that time based on the fact that no discrepancies were noted for the products being accepted, they were released to stock on the 18th july 2006.Failure analysis - the valve was visually inspected; biological debris was noted, as well as the stator and x ray dot were dislodged and the rickham connector was broken off (broken connector not returned).The position of the cam could not be determined as the stator was dislodged.The valve was hydrated.Due to dislodged stator the could not be tested for programming, reflux and pressure.Due to broken connector, the valve could not be flushed and tested for leak.Under magnification: a crack and a bump mark were noted in the valve casing.Corrosion was noted on the stator and the x ray dot.The root causes for the dislodged stator and x ray dot could be partly due to the valve receiving a hard knock.The root cause for the bump mark and crack in the valve casing is due to the valve receiving a hard knock.The root cause for the broken rickham connector could be due to the valve receiving a hard knock.The root cause of the corrosion could not be clearly determined.The potential cause of failure for the problem reported by the customer is probably due to the valve receiving a hard knock.
|