• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT DUO TUSM; SURGICAL LIGHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRUMPF MEDIZIN SYSTEME GMBH + CO. KG TRULIGHT DUO TUSM; SURGICAL LIGHT Back to Search Results
Model Number 1574850
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Patient Problem No Patient Involvement (2645)
Event Date 08/03/2020
Event Type  malfunction  
Manufacturer Narrative
A hillrom service technician inspected the surgical light system and found that the circlip was off the spring arm mount causing the spring arm to drop a few inches.The circlip was replaced, and the light is functioning as designed.A field corrective action has been initiated by trumpf medical and reported to the fda as a result of investigations into similar events (recall number z-563-2016).The investigations have found that improper installation or servicing of the circlip in this joint can result in the slipping down or falling of the spring arm and light head components.Based on this information, no further action is required.
 
Event Description
The customer alleged the trumpf monitor arm "c" bracket came loose causing monitor to drop down from support arm a few inches.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIGHT DUO TUSM
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld 07318
GM   07318
MDR Report Key10447494
MDR Text Key205019079
Report Number9681407-2020-00051
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 08/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number1574850
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/03/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-