A field service engineer (fse) was at the customer's site to address the reported event.Fse confirmed the flag from error log but could not confirm the incorrect patient results.Fse reproduced the hb-var problem by running the patient's samples on analyzer 1.While troubleshooting, fse found the retention time (rt) was running high, and the small syringe was leaking.Fse replaced the small syringe and the check valves, the rt dropped from 0.61 to 0.56minutes.Fse adjusted the flow rate to achieve rt of 0.59 minutes, then primed the analyzer but the hb-var persisted.Fse decontaminated the analyzer and ran the analyzer until the hb-var was gone.Fse successfully completed precision, calibration and qc runs without error and within acceptable range.No further action required by field service.The g8 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no similar complaints identified during the search period.The g8 variant operator's manual chapter 1, introduction and application states: interpretation of results: the sa1c measuring range is 4.0 - 16.9%.The ideal retention time for sa1c is 0.59 minutes.The ideal retention time for a0 is 0.90minutes.The g8 variant operator's manual chapter 3, provides detailed information on test report interpretation and analyzer flag settings.The most probable cause of the reported event is due to faulty small syringe and check valves and biological contamination of the system.
|