On (b)(6) 2020, the patient underwent a bronchoscopic lung volume reduction (blvr) procedure with zephyr valves.While attempting to insert the zephyr 4.0-j endobronchial delivery catheter (edc) into an airway through the bronchoscope while the bronchoscope was severely articulated, the edc met resistance and was eventually forced into the airway.Once the edc entered the airway, it was observed that a small piece of the housing on the distal end had broken away from the catheter.The edc was removed and they opted to use a new zephyr 4.0-j edc.The territory manager reminded the physician and support team to advance the edc to the end of the bronchoscope prior to articulating the bronchoscope.There were no further issues during the procedure.The broken fragment was not located or retrieved.
|