• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, B.V. (CD) SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, B.V. (CD) SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 25AGN-751
Device Problems Backflow (1064); Use of Device Problem (1670)
Patient Problem Aortic Regurgitation (1716)
Event Date 08/19/2020
Event Type  Injury  
Manufacturer Narrative
Further information regarding this event has been requested.Investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020 a double valve replacement and tv repair with laa ligation procedure was performed.The implanted 25agn-751 valve was found with transvalvular leakage under echo after procedure.No obstruction was observed during surgery.The 25agn-751 valve was explanted from the patient and replaced with 23agn.There was no transvalvular leakage found under echo.
 
Manufacturer Narrative
Additional information: d10, g4, h2, h3, h6 h10.The reported regurgitation could not be confirmed.No anomalies were found with the valve cusps, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, B.V. (CD)
lot 21
caguas west industrial park
caguas, puerto rico 00726 -099
MDR Report Key10461609
MDR Text Key204694859
Report Number3003681312-2020-00011
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Replace
Type of Report Initial,Followup
Report Date 11/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/14/2022
Device Model Number25AGN-751
Device Lot Number6137652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2020
Was the Report Sent to FDA? No
Date Manufacturer Received10/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-