Model Number EL5ML
Device Problems Failure to Form Staple (2579); Ejection Problem (4009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/05/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # u9459p.Investigation summary: the analysis results found that the el5ml device was received with one jaw disengaged from the cam.This condition would not allow the jaws to collapse in order to form the clips.No functional test was performed due the condition of the device.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.In addition, 3 clips were found inside clip track.Possible causes for the condition of the jaw disengaged from the cam may be inadvertent force, twisting or pressure being placed on the device jaws, using the jaws of the device as a dissector/retractor, or damage to the jaws while entering the trocar.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic cholecystectomy, the physician was doing the procedure when the 5ml clip applied broke after three clips.They opened a second applier and finished without incident.It was only closing the tip of the clip so it looked like a diamond shape wrapped around the vessel, but not occluded.It was ejecting clips before the jaw could squeeze it closed.Patient consequence was not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10463593
MDR Text Key204762252
Report Number3005075853-2020-04395
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Date Manufacturer Received08/05/2020
Date Device Manufactured05/28/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-