RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Hematoma (1884)
Event Date 08/11/2020
Event Type  Injury  
Manufacturer Narrative
Udi - (b)(4).Perforator was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
A physician reported the perforator failed to disengage in time during a craniotomy, punctured the dura mater, and caused a hematoma to the brain.A 20-minute delay in surgery was reported.
 
Manufacturer Narrative
The perforator was returned for evaluation: failure analysis: the perforator unit was inspected using the unaided eye.The perforator was visually soiled and had a slightly worn eto label.Ifu testing was performed with no observed anomalies.Functional testing was performed: the unit was found to perform as intended and fulfilled the acceptance criteria.In the failure analysis that was performed, the returned unit was found to work as intended, and met all acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10465903
MDR Text Key204746808
Report Number3014334038-2020-00027
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ6961T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/23/2020
Date Manufacturer Received11/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-