• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 LATEX FOLEY CATHETER Back to Search Results
Catalog Number UNKNOWN
Device Problems Burst Container or Vessel (1074); Material Rupture (1546); Device Fell (4014)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 05/26/2020
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the patient had indwelling catheter during the operation, the catheter fell out allegedly due to the balloon burst which resulted in postoperative complications.The catheter was replaced.
 
Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.The potential root cause for this failure mode could be user related (example: contact with sharp object)/ / exposure to petrolatum based products mechanical failure/operator error.The lot number was unknown; therefore, the device history record could not be reviewed.Unable to perform labelling review due to unknown product code.Although the product family was unknown, the latex foley catheter product ifus were found to be adequate based on past reviews.Correction: d4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event Description
It was reported that the patient had indwelling catheter during the operation, the catheter fell out allegedly due to the balloon burst which resulted in postoperative complications.The catheter was replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATEX FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10468980
MDR Text Key204861092
Report Number1018233-2020-05543
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 11/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received10/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-