• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKNOWN 44 +0 HEAD TRIAL; PROSTHSIS,HIP, HMI-,TRUNNIONBARING,FEMORAL, METAL/POLYACETAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKNOWN 44 +0 HEAD TRIAL; PROSTHSIS,HIP, HMI-,TRUNNIONBARING,FEMORAL, METAL/POLYACETAL Back to Search Results
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/05/2020
Event Type  Injury  
Event Description
It was reported that, during a thr surgery, the head trial broke while being used inside of the patient.Trialing was completed with the faulty device.Surgery was not delayed.The patient did not experience any other complication.
 
Manufacturer Narrative
The device, used in treatment was not returned for evaluation.Without the actual product, product evaluation could not be performed and complaint could not be confirmed.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.As device information was not made available, device history record , risk management review and complaint history review cannot be completed.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will not be issued for the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN 44 +0 HEAD TRIAL
Type of Device
PROSTHSIS,HIP, HMI-,TRUNNIONBARING,FEMORAL, METAL/POLYACETAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10471689
MDR Text Key204930787
Report Number1020279-2020-04296
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Date Manufacturer Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-