Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 08/24/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).(udi): n/a.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that a piece of the bolt cutter broke off during the procedure and the piece fell into the patient¿s wound.An x-ray was taken which revealed that the piece was inside the patient.The surgeon located the broken piece and removed it from the patient¿s wound.Attempts have been made and there is no additional information available at this time.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).Reported event was confirmed per visual inspection which shows one of the jaws fractured near the tip.X-rays were received however they were not sent for review as it would not enhance the investigation.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|