Model Number N/A
Device Problems Fitting Problem (2183); Connection Problem (2900)
Patient Problem No Patient Involvement (2645)
Event Date 04/24/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Current repair: product evaluation: product review of the air dermatome serial number (b)(4) by zimmer-(b)(6) on (b)(6) 2020 revealed that the handpiece and hose are difficult to connect due to scratches on the o-rings.The bearings were also deteriorated.Product repair: repair of the device was performed by zimmer-(b)(6) on (b)(6) 2020 which included replacement of the following: spring seal, needle bearing the device, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.Review of the device history records identified no deviations or anomalies during manufacturing.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that the dermatome was in need of repair during kit inspection because the connection is tight.During the investigation, it was revealed that the handpiece and hose are difficult to connect.There was no adverse event reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key10475915
MDR Text Key205123285
Report Number0001526350-2020-00738
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number62008594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/30/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-