• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER DUCT OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDA-004
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/30/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020 an amplatzer duct occluder was selected for a procedure.During deployment the device was not aligning with the cable.The physician made two attempt to bring back the device to its original shape but was unsuccessful.The screw was extending out of its normal shape.
 
Manufacturer Narrative
The reported event of device deformation upon deployment could not be confirmed.The occluder met functional specifications, deploying without deformity, when analyzed at abbott.The nitinol braid on one end of the device was damaged; consistent with damage during use.Two photos were received from the field, which appeared to show the delivery cable and end screw off of the center of the proximal end of the device and a deformed occluder.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER DUCT OCCLUDER
Type of Device
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key10476368
MDR Text Key205114333
Report Number2135147-2020-00386
Device Sequence Number1
Product Code MAE
UDI-Device Identifier00811806011110
UDI-Public00811806011110
Combination Product (y/n)N
PMA/PMN Number
P020024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-PDA-004
Device Catalogue Number9-PDA-004
Device Lot Number7122476
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2020
Was the Report Sent to FDA? No
Date Manufacturer Received11/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age2 YR
-
-