The reported event of device deformation upon deployment could not be confirmed.The occluder met functional specifications, deploying without deformity, when analyzed at abbott.The nitinol braid on one end of the device was damaged; consistent with damage during use.Two photos were received from the field, which appeared to show the delivery cable and end screw off of the center of the proximal end of the device and a deformed occluder.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|