Model Number 595000-001 |
Device Problems
Break (1069); Degraded (1153); No Visual Prompts/Feedback (4021)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The freedom driver will be evaluated by syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4) initial.
|
|
Event Description
|
The freedom driver was not supporting a patient.While performing a routine evaluation, the syncardia technician reported that the freedom driver exhibited a blank screen.
|
|
Manufacturer Narrative
|
Freedom driver was returned for service and failed incoming testing.Visual inspection of external components found the lcd display was cracked and darkened.Additional cosmetic damage to the display cover and front driver housing was observed.Visual inspection of internal components found scuff marks on the main pcba and primary, indicating contact occurred.The root cause of the reported issue could not be determined as no additional information was provided from the hospital returning the freedom driver for service.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.(b)(4) follow-up report 1.
|
|