Model Number 595000-001
Device Problems Break (1069); Degraded (1153); No Visual Prompts/Feedback (4021)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/05/2020
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be evaluated by syncardia.The results of the investigation will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The freedom driver was not supporting a patient.While performing a routine evaluation, the syncardia technician reported that the freedom driver exhibited a blank screen.
 
Manufacturer Narrative
Freedom driver was returned for service and failed incoming testing.Visual inspection of external components found the lcd display was cracked and darkened.Additional cosmetic damage to the display cover and front driver housing was observed.Visual inspection of internal components found scuff marks on the main pcba and primary, indicating contact occurred.The root cause of the reported issue could not be determined as no additional information was provided from the hospital returning the freedom driver for service.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.If new or additional information is received in the future, syncardia will file a follow-up mdr.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10477307
MDR Text Key205470078
Report Number3003761017-2020-00171
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/05/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-