• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK FEMORAL-RM-LL-SZ 3; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK FEMORAL-RM-LL-SZ 3; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Model Number 180513
Device Problem Loss of Osseointegration (2408)
Patient Problem Injury (2348)
Event Date 08/06/2020
Event Type  Injury  
Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Device not returned.
 
Event Description
The doctor revised a mako mck medial pkr due to aseptic loosening of the femoral component.He thinks it was due to cement technique at time of implantation.The construct is two years old.There is visible wear on the poly insert.He believes it is too young to have experienced poly wear, and would like the insert to be examined, assessed and reported back to him please.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK FEMORAL-RM-LL-SZ 3
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
alessandra chavez
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key10477460
MDR Text Key205130661
Report Number0002249697-2020-01774
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier00848486000622
UDI-Public00848486000622
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number180513
Device Catalogue Number180513
Device Lot Number542190-M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age70 YR
Patient Weight91
-
-