This event has been recorded by zimmer biomet under (b)(4).Concomitant medical products: item# 00880100200, dermatome hose, lot# 555350.Air dermatome handpiece :product review of the zimmer air dermatome serial number (b)(4) by a zimmer biomet certified service repair technician on 28 february 2020 revealed that the unit¿s rpms were not in specification, the control bar was not in the correct position and the calibration was out at the 0 reading.Repair of the device was performed by a zimmer biomet certified service repair technician on 28 february 2020.The device, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Devices are used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|