Model Number N/A
Device Problems Power Problem (3010); Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Date 02/13/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Concomitant medical products: item# 00880100200, dermatome hose, lot# 555350.Air dermatome handpiece :product review of the zimmer air dermatome serial number (b)(4) by a zimmer biomet certified service repair technician on 28 february 2020 revealed that the unit¿s rpms were not in specification, the control bar was not in the correct position and the calibration was out at the 0 reading.Repair of the device was performed by a zimmer biomet certified service repair technician on 28 february 2020.The device, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Devices are used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that during case set-up the handpiece didn't allow proper flow of nitrogen through it for power.No harm or delay was reported.Investigation showed the rpm's were low and outside of specifications.There was no adverse event reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key10478592
MDR Text Key205473006
Report Number0001526350-2020-00742
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024375901
UDI-Public(01)00889024375901(11)160623(10)63394127
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 09/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number63394127
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/21/2020
Was the Report Sent to FDA? No
Date Manufacturer Received09/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-