(b)(4).The returned ultratome xl was analyzed, and a visual evaluation noted that the working length was twisted and the distal pierced hole was torn which displaced the cutting wire anchor from it's original position.A functional evaluation noted that a test guidewire could not pass through the distal tip due to the wire anchor being dislodged.Additionally, water flowed through the injection port as intended and the resistance across the 2-in-1 connector and cutting wire were within specification.No other issues with the device were noted.The reported event was confirmed.According to the product analysis, the distal pierce hole was torn and the working length twisted causing the cutting wire anchor to displace and tear the catheter.Additionally, a test guidewire was not able to pass through the distal tip due to the wire anchor being dislodged.Based on the condition of the device the failure found could had been generated due to factors encountered during the procedure, the technique used, the patient anatomy and the interaction with the scope, also if the device was actuated in coil position.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu) / product label.
|