SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTH, HIP, SEMI-CONSTRA, UNCEMENT, METAL/POLY, NON-POROUS, CALICUM-PHOSPHATE
|
Back to Search Results |
|
Model Number 71360093 |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/05/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that the femoral head/neck impactor was found to be broken during a routine efip inspection.There were no reported problems from any cases.
|
|
Manufacturer Narrative
|
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirms the femoral head is missing from the device.The femoral head was not returned with the device.This device shows signs of significant wear/usage.This device was manufactured in 2012.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
|
|