The wrong analysis results were sent with follow up 2.The correct analysis results are now attached.The returned icd was visually inspected after receipt, and signs of abrasion were found at the housing.The initial interrogation confirmed the clinical observation, the device could no longer be interrogated.In the next step, the icd was opened, and the internal structure was examined.Damage to the fuse that separates the battery from the electronic module and to the final shock stages was detected.These damages are with high probability the result of a shock delivery into an external low-ohmic shock path.Due to the damage to the module, the icd could no longer be interrogated.Possible clinical complications that might lead to an external short-circuit are, among others, the twiddler syndrome, the subclavian crush syndrome, or other damage to the lead insulation, which leads to a low-ohmic contact of the high-voltage conductors.The manufacturing process of this device was reviewed.The production documents showed no anomalies.All manufacturing steps had been carried out correctly.A performed standard final electrical test documented proper device behavior.There were no indications of a material defect or manufacturing error.
|