Model Number 682000 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
|
Event Description
|
The account alleges that during the hemodynamic monitoring of a patient, the pressure monitoring pm set was found to be leaking fluid from the 3-way stopcock.The pm set was exchanged, and the procedure was successfully completed with no consequences to the patient.
|
|
Manufacturer Narrative
|
The suspect device has been returned for evaluation.Functional test was conducted using saline set-up pressurized at 300mmhg and leakage was found from the planecta stopcock housing.The complaint is confirmed.The root cause could not be determined however, the cracked propagation on the housing could be likely due to force or weight applied onto the planecta resulted to mechanical stress outside and incurred during procedure.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.
|
|
Search Alerts/Recalls
|