Catalog Number 03666794001 |
Device Problems
Circuit Failure (1089); Display Difficult to Read (1181)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Occupation is lay user/patient.The customer¿s device was requested for investigation, but has not been returned.The investigation is ongoing.
|
|
Event Description
|
The initial reporter complained of a cc-xs meter display issue, which could cause misinterpretation of results.The reporter stated the meter has provided unknown errors and power issues.The reporter performed a display test, and segments were missing from the results field and date field.Since the meter display issue was discovered, the patient has only used the meter two times for testing.The patient has been getting blood drawn for laboratory testing.No misinterpretation of results or serious injuries were reported.
|
|
Manufacturer Narrative
|
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found the printed circuit board was contaminated.The observed contamination can temporarily lead to the reported display issue.The root cause is determined to be contamination of the contacts due to improper handling or maintenance by the customer.Medwatch fields d9 and h3 were updated.
|
|