MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE SIERRA; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550350-28
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Patient Involvement (2645)
Event Date 08/14/2020
Event Type  malfunction  
Manufacturer Narrative
The product was not returned to abbott vascular for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints reported from this lot.The investigation was unable to determine a conclusive cause for the reported packaging damage and sterility issue.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that during unpacking of a 3.50x28mm xience sierra, the foil pouch was noted without any damages, however, the inside clear tyvek pouch was noted with a hole in it.There were no damages noted to the chipboard box.The device was not used and there was no patient involvement.Another device was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA
Type of Device
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10489198
MDR Text Key212542983
Report Number2024168-2020-07363
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648227448
UDI-Public08717648227448
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/15/2021
Device Model Number1550350-28
Device Catalogue Number1550350-28
Device Lot Number0060242
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-