It was reported that on (b)(6) 2020 a (b)(6) female patient underwent an on-pump heparinized mitral valve repair, maze, left atrial appendage management procedure.During the procedure the surgeon was using a re-usable retractor device, as he tightened the knob to lock the retractor arm in place, the internal wire came loose from the assembly and the arm lost its rigidity.The surgeon then attempted to remove the retractor and the wire and segments came free from the assembly and several of the segments slipped off the wire and fell into the surgical site.The surgeon retrieved the segments that were immediately visible, and a back-up retractor arm was used then to complete the procedure.Post-surgery patient had an x-ray study which confirmed that no segments were left behind inside the patient.The procedure was prolonged for 20min, there was no injury to the patient.
|