|
Catalog Number 05265231001 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Local fse replaced the damaged components and the instrument is running as intended.The affected parts were discarded at the customer site.The investigation is ongoing.(b)(4).
|
|
Event Description
|
An (b)(6) customer reported smoke and a burning smell coming from the benchmark xt instrument.The local field service engineer (fse) found that a fitting broke within the instrument, resulting in leaking liquids, which caused an electrical short.Consequently, the transformer burnt up.The customer opened the window and left the lab.There was no report of harm or injury.
|
|
Manufacturer Narrative
|
Investigation found a broken quick disconnect fitting allowed reagent leakage onto the power transformer.The power transformer experienced a temporary low current shorting due to reagent ingress which caused insulation charring/smoke which then later self-extinguished once the reagent evaporated.The system is (b)(4) certified.The (b)(4) certification evaluates materials of construction for flammability and provides independent assurance that any internal fire would not or would be unlikely to spread outside the instrument frame.Additionally, (b)(4) evaluates that thermal and electrical fuses cut power in runaway conditions.The capa investigation of the fluids leak has identified several contributing factors, including but not limited to the impact of material compatibility on potential leak sources and implementation of several categories of suboptimal leak mitigation measures within the instrument (for example, lack of hose clamps to keep fluidics tubing from separating from fittings).Improvements have been made to the ventana design process, including design readiness and material compatibility work instructions and checklists.(b)(4).
|
|
Search Alerts/Recalls
|
|
|