• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Pressure Problem (3012)
Patient Problem No Patient Involvement (2645)
Event Date 08/24/2020
Event Type  malfunction  
Manufacturer Narrative
Vyaire file identification: (b)(4).Any additional information received from the customer will be included in a follow-up report.At this time, the suspect device has not been return for investigation.Therefore, no further evaluation can be identified.
 
Event Description
The customer reported that a 3100a ventilator will not pressurize.No patient involvement.The reported issue was detected prior to use.
 
Manufacturer Narrative
Result of investigation: : a vyaire field service representative (fsr) went onsite and evaluated the ventilator.Fsr found that the adjust knob is loose.Realigned adjust knob and do minor adjustment on pr3.Performed pt cal and performance test, both passed per 3100a label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, ca
MDR Report Key10498341
MDR Text Key205905267
Report Number2021710-2020-12501
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/24/2020
Initial Date FDA Received09/06/2020
Supplement Dates Manufacturer Received09/25/2020
Supplement Dates FDA Received10/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-