Model Number 3100 A |
Device Problem
Pressure Problem (3012)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Vyaire file identification: (b)(4).Any additional information received from the customer will be included in a follow-up report.At this time, the suspect device has not been return for investigation.Therefore, no further evaluation can be identified.
|
|
Event Description
|
The customer reported that a 3100a ventilator will not pressurize.No patient involvement.The reported issue was detected prior to use.
|
|
Manufacturer Narrative
|
Result of investigation: : a vyaire field service representative (fsr) went onsite and evaluated the ventilator.Fsr found that the adjust knob is loose.Realigned adjust knob and do minor adjustment on pr3.Performed pt cal and performance test, both passed per 3100a label.
|
|