• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI INC SHOCKPULSE-SE LITHOTRIPSY SYSTEM (REUSAB; LITHOTRIPTOR, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI INC SHOCKPULSE-SE LITHOTRIPSY SYSTEM (REUSAB; LITHOTRIPTOR, ULTRASONIC Back to Search Results
Model Number SPL-SR
Device Problems Defective Device (2588); Output Problem (3005)
Patient Problem No Patient Involvement (2645)
Event Date 07/30/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device has not yet been received for evaluation.The cause of the issue cannot be determined at this time.If additional information becomes available this report will be supplemented accordingly.
 
Event Description
It was reported that the device transducer receptor malfunctioned.The issue found during incoming inspection.According to the reporter, the device when press switch on transducer handling ,¿high mode/standard mode¿ had switched.There was no patient involvement on this report, no user harm or injury reported due to the event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide review of the device history records (dhr), unique device identifier (udi) number and investigation conclusion.Please see updated sections: d4,d10,g4, g7, h2, h3,h4, h6 and h10.The device was evaluated by olympus thailand.The user report was confirmed.It was found that transducer receptacle damage resulted in the reported phenomenon.Device history records were reviewed and showed the product met all specifications upon release.Damage to the transducer receptacle is often incurred as a result of user error, often the user does not realize all connections are push/pull and instead the plug is twisted upon connection or removal.This action results in damage to the pins internal to the socket and may crack or damage the housing.As stated in the instruction for use: connect the transducer to the generator by pressing the plug straight in.Caution do not twist or turn the plug.Connect the transducer to the generator by aligning the key way of the transducer connector with the key way slot on the transducer receptacle on the front panel.Push straight in.Caution do not twist or turn the plug.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE-SE LITHOTRIPSY SYSTEM (REUSAB
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI INC
136 turnpike road
southborough PA 01772
MDR Report Key10501483
MDR Text Key206822077
Report Number3011050570-2020-00066
Device Sequence Number1
Product Code FEO
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-SR
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/10/2020
Initial Date FDA Received09/08/2020
Supplement Dates Manufacturer Received01/11/2021
Supplement Dates FDA Received01/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-