• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Model Number 500101-001
Device Problem Mechanical Problem (1384)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
The cpc connector will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The cpc (colder products company) connector is a component that provides the interface between the driver drivelines and the tah-t cannula.The customer, a syncardia certified hospital, reported that the spring in the cpc connector was found to be sideways.The doctor pulled the spring out with a hemostat, removed the connector and replaced it with a new connector.There was no reported adverse patient impact.
 
Manufacturer Narrative
Syncardia has a corrective and preventive action (capa) to address the issue of missing or displaced springs in cpc connectors.From the results of the capa investigation, the primary root cause of spring displacement within the cpc connectors is the use of a wire tie within the cpc connector body (which can come in contact with the spring).Clinicians and patients are trained to thread a wire tie beneath the thumb release tab of the cpc connectors to prevent accidental disconnection of the cannulae from the drivelines.It is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch out.Syncardia capa (b)(4), cpc connectors, is currently at step 5 action plan.Syncardia has completed its evaluation and is closing this file.Ce 5354 follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key10503229
MDR Text Key205954148
Report Number3003761017-2020-00189
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 02/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number110942
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2021
Date Manufacturer Received08/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age29 YR
-
-