Model Number 500101-001 |
Device Problems
Material Discolored (1170); Material Too Rigid or Stiff (1544)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A piece of the darkened cannulae will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce (b)(4) initial.
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the tah-t cannula were discolored, "super dark:" all the way up to the patient's exit site.The customer also reported that the patient is a heavy smoker.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The cannula piece was sent to an outside laboratory, metl, for analysis via ftir-atr per astm e 1252.The report attributed the discoloration to photodegradation (likely due to the sun), which can result in a brown discoloration and more brittle properties.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.Ce 5355 (b)(4) follow-up report 1.
|
|