• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T); BIVENTRICULAR REPLACEMENT DEVICE Back to Search Results
Model Number 500101-001
Device Problems Material Discolored (1170); Material Too Rigid or Stiff (1544)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/26/2020
Event Type  malfunction  
Manufacturer Narrative
A piece of the darkened cannulae will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce (b)(4) initial.
 
Event Description
The customer, a syncardia certified hospital, reported that the tah-t cannula were discolored, "super dark:" all the way up to the patient's exit site.The customer also reported that the patient is a heavy smoker.There was no reported adverse patient impact.
 
Manufacturer Narrative
The cannula piece was sent to an outside laboratory, metl, for analysis via ftir-atr per astm e 1252.The report attributed the discoloration to photodegradation (likely due to the sun), which can result in a brown discoloration and more brittle properties.This issue will be tracked and trended as part of the customer complaint process.Syncardia has completed its investigation and is closing this file.Ce 5355 (b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TEMPORARY TOTAL ARTIFICIAL HEART (TAH-T)
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson, az
Manufacturer Contact
sagar pimpalwar
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key10503232
MDR Text Key206129931
Report Number3003761017-2020-00190
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number110942
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age29 YR
Patient SexMale
-
-